Abstract Number: PB1108
Meeting: ISTH 2022 Congress
Background: The thrombin generation assay (TGA) is a global haemostatic test widely utilized to investigate both diseases and drugs that cause thrombosis and bleeding.
Aims: To establish the reference intervals for TGA in healthy Asian adults and measure the variability of 3 standardised tests: STG-BleedScreen, STG-DrugScreen and STG-ThromboScreen.
Methods: Whole blood was obtained from 125 healthy Asian adults aged 18 to 65 years (Chinese (71.2%, n=89), Indians (8%, n=10), Malays (15.2%, n=19), Filipinos (4.8%, n=6) and others (0.8%, n=1)). 51.20% (n=65) were women. TGA was performed on platelet poor plasma using the ST-Genesia (STG) (Stago, Asnières-sur-Seine, France). 125 samples were assayed using the STG-BleedScreen and STG-ThromboScreen, with 62 samples measured using the STG-DrugScreen. Subgroup analysis of sex and race was performed.
Results: Reference ranges were established with both normalized and absolute TG parameters (Table 1, Table 2). STG bleed screen (normalized) had a lag time ratio 1.07±0.2, peak height 103.72±31.4%, time to peak ratio 1.12±0.1, ETP 101.59±22.1%, Vel. Index 95.90±37.5%, start tail ratio 0.99±0.1 min. STG drug screen (normalized) had a lag time ratio 1.00±0.17, peak height 96.52±9.29%, time to peak ratio 1.01±0.12, ETP 88.98±12.72%, Vel. Index 99.51±18.35%, start tail ratio 0.90±0.08 min. STG Thromboscreen (without thrombomodulin) (normalized) had a lag time ratio 1.11±0.19, peak height 84.19±20.14%, time to peak ratio 1.23±0.17, ETP 93.51±15.06%, Vel. Index 73.30±30.09%, start tail ratio 1.08±0.15 min. STG Thromboscreen (with thrombomodulin) (absolute) had a lag time 2.42±0.47 min, peak height 101.57±46.51%, time to peak 4.31±0.63 min, ETP 427.75±189.04 nM.min, ETP Inhibition 64.85±13.22%, Vel. Index 77.19±40.66 nm/min, start tail 13.63±0.97 min. The coefficient of variation for ETP was lower than the reported 23% and the ETP was decreased by 65% when TM was added. Intergender and interethnic variation were not statistically significant.
Conclusion(s): STG appears to be suitable for the accurate measurement of TG in healthy Asian adults.
Table
Tables 1 and 2: Interindividual variability and reference intervals of thrombin generation parameters measured by ST Genesia in healthy Asian adults using the STG-BleedScreen, STG-DrugScreen and STG-ThromboScreen assays.
To cite this abstract in AMA style:
Fan B, Ang M, Tan H, Boulaire J, Chai T, Ong K, Sampath V, SUM C. Thrombin Generation Using the ST-Genesia System in a Cohort of Healthy Asian Adults: Absolute and Normalized Values [abstract]. https://abstracts.isth.org/abstract/thrombin-generation-using-the-st-genesia-system-in-a-cohort-of-healthy-asian-adults-absolute-and-normalized-values/. Accessed September 27, 2023.« Back to ISTH 2022 Congress